A new instrument for surgical stabilization of multiple rib fractures
Xiong Jian1, Wu Lei1, Pi Yuyang1
1Department of Thoracic Surgery, Shanghai Pudong Hospital, Fudan University Pudong Medical Center, Pudong, Shanghai, China.
Objective:
Rib fixation is an effective treatment for patients with multiple rib fractures. We retrospectively evaluated the application of a four-claw titanium plate in patients with rib fractures.
Methods:
Fifty-four patients treated for multiple rib fractures in our hospital from 2012 to 2016 were divided into a surgery group (n = 27) and conservative treatment group (n = 27). The patients' age, sex, cause of fracture, Injury Severity Score, chest Abbreviated Injury Scale score, number of ventilator days, and length of hospitalization were recorded.
Results:
The mean duration of mechanical ventilation was 4.5 ± 0.7 and 7.9 ± 1.7 days in the surgery and control group, respectively, with a significant difference. The length of intensive care unit stay was also significantly different between the groups (5.9 ± 0.6 vs. 10.6 ± 1.9 days, respectively). The length of hospital stay and recovery time to regular life in the surgery and control group were 11.5 ± 1.9 and 3.9 ± 4.0 days and 38.2 ± 8.3 and 60.8 ± 12.1 days, respectively, both with significant differences.
Conclusion:
A four-claw titanium plate is valuable for patients with multiple rib fractures, allowing easy fixation of broken ribs beneath the scapula, even the second rib.
More Related Videos
Related Concept Videos
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
The Thoracic Cage: Ribs
Parts of a Typical Rib
A typical rib has a head, neck, and body. The posterior end of the rib is called the head, followed by a narrow neck. The head articulates primarily with the costal...