Related Experiment Video
Updated: Jan 18, 2026

Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Very complex internal standard response variation in LC-MS/MS bioanalysis: root cause analysis and impact assessment
Nico C van de Merbel1,2, Remco A Koster1, Corey Ohnmacht3
1Bioanalytical Laboratory, PRA Health Sciences, Amerikaweg 18, 9407 TK, Assen, The Netherlands.
Internal standards (ISs) are crucial for accurate quantitative bioanalytical LC-MS/MS methods. This study explores complex IS response variations, offering a decision tree to identify root causes like injection errors or sample degradation.
Area of Science:
- Analytical Chemistry
- Bioanalysis
- Mass Spectrometry
Background:
- Internal standards (ISs) are vital for quantitative bioanalytical methods.
- ISs correct for analytical response variations due to experimental conditions.
- Variability in IS response can indicate issues in sample handling or analysis.
Purpose of the Study:
- To investigate complex variations in internal standard (IS) responses.
- To identify analytical and matrix-related root causes of extreme IS variability.
- To propose a decision tree for diagnosing IS variability issues.
Main Methods:
- Case study analysis of complex IS response variations.
- Identification of analytical problems (e.g., injection errors, sample inhomogeneity).
- Assessment of matrix-related issues (e.g., degradation, ionization efficiency).
Main Results:
- Complex IS variability can stem from injection errors, sample inhomogeneity, degradation, and increased ionization efficiency.
- These issues significantly impact the reliability of quantitative bioanalytical methods.
- A systematic approach is needed to diagnose the source of variability.
Conclusions:
- Extreme internal standard variability in LC-MS/MS methods signals underlying analytical or matrix problems.
- A structured diagnostic approach, like the proposed decision tree, is essential.
- Addressing these root causes ensures the development of robust and reliable bioanalytical methods.
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...