Related Experiment Video
Updated: Jan 18, 2026

07:00
Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
1.4K
Recommendations for Measuring the Electrically Evoked Compound Action Potential in Children With Cochlear Nerve
Shuman He1,2, Xiuhua Chao3, Ruijie Wang3
1Department of Otolaryngology - Head and Neck Surgery, The Ohio State University, Columbus, Ohio, USA.
Ear and Hearing
|October 1, 2019
Summary
This study presents a new method for measuring electrically evoked compound action potentials (eCAP) in children with cochlear nerve deficiency (CND). The recommended parameters improve eCAP measurement success rates in these patients.
Area of Science:
- Audiology
- Neuroscience
- Biomedical Engineering
Background:
- Cochlear nerve deficiency (CND) presents unique challenges for auditory rehabilitation.
- Accurate measurement of electrically evoked compound action potentials (eCAP) is crucial for programming cochlear implants.
Purpose of the Study:
- To report a refined method for measuring eCAP in children with CND.
- To improve the success rate of eCAP measurements in this specific pediatric population.
Main Methods:
- Development of a three-step method: initial screen, pulse phase duration optimization, and eCAP threshold determination.
- Utilized recommended stimulating and recording parameters based on experience with 50 children with CND and Cochlear Nucleus implants.
- Compared results with manufacturer-default parameters.
Main Results:
- The developed method demonstrated a higher success rate for eCAP measurement in children with CND compared to default parameters.
- Successful eCAP measurement was achieved using the recommended parameters.
Conclusions:
- The proposed method is effective for measuring eCAP in children with CND in clinical settings.
- The method is not suitable for intraoperative recordings due to the time required.