Improvement of image quality applying iterative scatter correction for grid-less skeletal radiography in trauma room
Christoph G Lisson1, Catharina S Lisson1, Daniel Vogele1
1Department of Diagnostic and Interventional Radiology, Ulm University Medical Center, Ulm, Germany.
Background:
Iterative reconstruction is well established for CT. Plain radiography also takes advantage of iterative algorithms to reduce scatter radiation and improve image quality. First applications have been described for bedside chest X-ray. A recent experimental approach also provided proof of principle for skeletal imaging.
Purpose:
To examine clinical applicability of iterative scatter correction for skeletal imaging in the trauma setting.
Material And Methods:
In this retrospective single-center study, 209 grid-less radiographs were routinely acquired in the trauma room for 12 months, with imaging of the chest (n = 31), knee (n = 111), pelvis (n = 14), shoulder (n = 24), and other regions close to the trunk (n = 29). Radiographs were postprocessed with iterative scatter correction, doubling the number of images. The radiographs were then independently evaluated by three radiologists and three surgeons. A five-step rating scale and visual grading characteristics analysis were used. The area under the VGC curve (AUCVGC) quantified differences in image quality.
Results:
Images with iterative scatter correction were generally rated significantly better (AUCVGC = 0.59, P < 0.01). This included both radiologists (AUCVGC = 0.61, P < 0.01) and surgeons (AUCVGC = 0.56, P < 0.01). The image-improving effect was significant for all body regions; in detail: chest (AUCVGC = 0.64, P < 0.01), knee (AUCVGC = 0.61, P < 0.01), pelvis (AUCVGC = 0.60, P = 0.01), shoulder (AUCVGC = 0.59, P = 0.02), and others close to the trunk (AUCVGC = 0.59, P < 0.01).
Conclusion:
Iterative scatter correction improves the image quality of grid-less skeletal radiography in the clinical setting for a wide range of body regions. Therefore, iterative scatter correction may be the future method of choice for free exposure imaging when an anti-scatter grid is omitted due to high risk of tube-detector misalignment.
More Related Videos
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...