Comprehensive characterization and resolution of discrepant spectrophotometric bilirubin results in patients on
Timothy H T Cheng1, Teresa K C Tsui1, Jeffrey S S Kwok1
1Department of Chemical Pathology, Prince of Wales Hospital, The Chinese University of Hong Kong, Shatin, Hong Kong.
Background Eltrombopag is a thrombopoietin receptor agonist used for the treatment of thrombocytopenic conditions. It can cause pH-dependent discoloration of plasma/serum. Eltrombopag is potentially hepatotoxic. It can affect the assessment of hyperbilirubinemia because of its (i) absorbance at ~450 nm (bilirubin), (ii) absorbance at ~550 nm (diazo-bilirubin) and (iii) it can cause yellowish discoloration of the eyes at normal circulating bilirubin levels. Methods We collected 66 samples from patients on a range of eltrombopag dosages up to 150 mg daily. Bilirubin was measured using multiple routine spectrophotometric analyzers, the Doumas reference method and high-performance liquid chromatography (HPLC). Plasma/serum eltrombopag concentrations were determined using liquid chromatography tandem mass spectrometry (LC-MS/MS). Spike-in and admixture experiments delineated the effects of eltrombopag and its metabolites. Results Forty-nine of 52 samples from patients on ≥50 mg daily eltrombopag therapy showed significantly discrepant inter-analyzer total bilirubin results, a difference up to 64 μmol/L (3.7 mg/dL). In one sample, total bilirubin varied from 8 to 65 μmol/L (0.4-3.8 mg/dL) by different routine analyzers, with direct bilirubin ≤4 μmol/L (0.2 mg/dL). There was a positive correlation between total bilirubin difference and plasma eltrombopag concentration (r = 0.679), and spike-in experiments demonstrated that Beckman AU and Doumas reference methods were susceptible to positive interference. HPLC can quantify bilirubin after separating eltrombopag, and results suggest different analyzers are affected to varying degrees by eltrombopag and its metabolites. Conclusions Eltrombopag and its metabolites can cause positive interference to the spectrophotometric measurements of total bilirubin. Accurate measurements of total bilirubin may improve our understanding of the prevalence of hyperbilirubinemia in patients on eltrombopag therapy.
Background Eltrombopag is a thrombopoietin receptor agonist used for the treatment of thrombocytopenic conditions. It can cause pH-dependent discoloration of plasma/serum. Eltrombopag is potentially hepatotoxic. It can affect the assessment of hyperbilirubinemia because of its (i) absorbance at ~450 nm (bilirubin), (ii) absorbance at ~550 nm (diazo-bilirubin) and (iii) it can cause yellowish discoloration of the eyes at normal circulating bilirubin levels. Methods We collected 66 samples from patients on a range of eltrombopag dosages up to 150 mg daily. Bilirubin was measured using multiple routine spectrophotometric analyzers, the Doumas reference method and high-performance liquid chromatography (HPLC). Plasma/serum eltrombopag concentrations were determined using liquid chromatography tandem mass spectrometry (LC-MS/MS). Spike-in and admixture experiments delineated the effects of eltrombopag and its metabolites. Results Forty-nine of 52 samples from patients on ≥50 mg daily eltrombopag therapy showed significantly discrepant inter-analyzer total bilirubin results, a difference up to 64 μmol/L (3.7 mg/dL). In one sample, total bilirubin varied from 8 to 65 μmol/L (0.4-3.8 mg/dL) by different routine analyzers, with direct bilirubin ≤4 μmol/L (0.2 mg/dL). There was a positive correlation between total bilirubin difference and plasma eltrombopag concentration (r = 0.679), and spike-in experiments demonstrated that Beckman AU and Doumas reference methods were susceptible to positive interference. HPLC can quantify bilirubin after separating eltrombopag, and results suggest different analyzers are affected to varying degrees by eltrombopag and its metabolites. Conclusions Eltrombopag and its metabolites can cause positive interference to the spectrophotometric measurements of total bilirubin. Accurate measurements of total bilirubin may improve our understanding of the prevalence of hyperbilirubinemia in patients on eltrombopag therapy.
More Related Videos
08:42Using Multi-fluorinated Bile Acids and In Vivo Magnetic Resonance Imaging to Measure Bile Acid Transport
Published on: November 27, 2016
07:44Endoscopic Ultrasound-Guided Biliary Drainage: Endoscopic Ultrasound-Guided Hepaticogastrostomy in Malignant Biliary Obstruction
Published on: March 25, 2022
Related Concept Videos
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Effect of Hepatic Disease on Pharmacokinetics: Drug Dosing and Hepatic Blood Flow
Effect of Hepatic Disease on Pharmacokinetics: Dose Adjustments Due to Hepatic Impairment
Effect of Hepatic Disease on Pharmacokinetics: Active Drug, Metabolite and Fraction of Metabolized Drug