Related Experiment Video
Updated: Jan 6, 2026

Preparation and Evaluation of 99mTc-labeled Tridentate Chelates for Pre-targeting Using Bioorthogonal Chemistry
Published on: February 4, 2017
Precise measurement of intradermal fluid delivery using a low activity technetium-99m pertechnetate tracer
Sahan A Ranamukhaarachchi1, Tullio V Esposito2, Mehrsa Raeiszadeh3
1Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, BC V6T 1Z4, Canada; Faculty of Pharmaceutical Sciences, University of British Columbia, Vancouver, BC V6T 1Z3, Canada.
Abstract:
A method was developed and validated to determine the intradermal (ID) fluid delivery potential of several ID devices, including hollow microneedles. The novel method used water soluble technetium-99 m pertechnetate (99mTcO4-) diluted in normal saline to measure the volume of fluid delivered to and remaining in the skin. The fluid that back-flowed to the skin surface and the fluid left on the device surface were also quantified, thus capturing all fluid volumes deposited during intradermal injections. The technique described in this manuscript was used to assess the injection performance of conventional hypodermic needles and hollow microneedles ex vivo using porcine skin and in vivo with a rat model. Since only a small fraction, 1.1%, of the water-soluble tracer remained bound to the skin when applied topically, the technique can be used to differentiate between injected fluid and backflow. Counting of gamma radiation from 99mTcO4- provided sub-nanoliter resolution for volume measurements, making the proposed method powerful, sensitive, and suitable for the assessments of ID injection devices, particularly for vaccine delivery.