Implementation and Evaluation of Phantomless Intensity Modulated Radiotherapy Delivery Verification Using
Buchapudi Rekha Reddy1, Manickam Ravikumar2, Chandraraj Varatharaj1
1Dept. of Radiation Physics, Kidwai Memorial Institute of Oncology, Bangalore, India.
FractionCHECKTM software efficiently verifies dynamic IMRT delivery accuracy using DynaLog files without external devices. This tool provides a fast, reliable method for quality assurance in busy radiotherapy departments.
Area of Science:
- Medical Physics
- Radiotherapy
- Quality Assurance
Background:
- Dynamic Intensity-Modulated Radiation Therapy (IMRT) requires precise delivery verification.
- Traditional verification methods can be time-consuming and require external measurement devices.
- Automated software solutions are needed to improve efficiency and accuracy in IMRT QA.
Purpose of the Study:
- To implement and evaluate the DynaLog file-based FractionCHECKTM software for automated IMRT delivery verification.
- To assess the software's performance in analyzing treatment accuracy for various patient cases.
- To determine the pass/warn/fail rates and analyze errors in dynamic IMRT deliveries.
Main Methods:
- Analysis of DynaLog files from 40 dynamic IMRT cases across different treatment sites (head & neck, brain, cervix, esophagus).
- Validation of FractionCHECKTM software using standard DMLC QA test fields.
- Calculation of mean RMS error, 95th percentile leaf error, and gamma analysis (1%, 1mm criterion) for planned vs. delivered fluences.
Main Results:
- Analysis of 1134 fractions and 25,166 log files from 40 patients.
- Overall pass rate of 95.4% (23,906/25,166 files), with 4.1% warning (1022) and 0.9% failure (238).
- Mean RMS error of 0.34±0.08 mm and mean 95th percentile leaf error of 0.66 ± 0.18 mm, meeting established safety standards.
Conclusions:
- FractionCHECKTM software offers an efficient, automated method for verifying IMRT treatment deliveries.
- The tool can be rapidly implemented, minimizing delays in busy clinical settings.
- FractionCHECKTM is a powerful, simple tool for routine QA and treatment delivery assessment in IMRT.