Dual energy computed tomography analysis in cancer patients: What factors affect iodine concentration in contrast
Giuseppe Corrias1, Peter Sawan2, Usman Mahmood3
1Department of Radiology, Memorial Sloan Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA; Department of Radiology, University of Cagliari, Via Università, 40, 09124, Cagliari, CA, Italy.
Purpose:
The aim of the study is to explore the patient's and scan's parameters that affect the iodine concentration in the abdomen using dual energy computed tomography (DECT) in an oncologic population.
Method:
This is a retrospective study with consecutive patients with different cancers who underwent a single-source DECT (ssDECT) examinations at our institution between years 2015 and 2017. On axial IODINE images, the radiologist manually drew a circular ROI along the inner contour of the aorta. Mean iodine concentration and ROI areas were recorded. Body mass index for every patient was recorded. Descriptive statistics were summarized for iodine concentration and patient/scan characteristics. Linear regression was used to examine associations between iodine concentration in aorta and studied characteristics. Statistical significance was set at a p value < 0.05.
Results:
The univariate analysis, showed a statistically significant association between iodine concentration within the aorta and the area of ROI (Estimated Coefficient β: -0.013), the rate of injection (Estimated Coefficient β: 2.09), the acquisition time (Estimated Coefficient β: -0.195). In multivariable analysis iodine concentration in the aorta increased with higher rate of injection (4 ml/sec), smaller ROI area and lower BMI.
Conclusion:
Our results showed how iodine concentration is highly dependent on some intrinsic and extrinsic parameters of the examination. These parameters should be taken into account since lower concentration of iodine decrease contrast-to-noise ratio, and in longitudinal follow up studies, they would affect iodine quantitive assessments in cancer patients with frequent chemotherapy-induced variations in BMI and cardiac function.
More Related Videos
06:52Positron Emission Tomography Using 64-Copper as a Tracer for the Study of Copper-Related Disorders
Published on: April 28, 2023
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...