Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
Interleaved, undersampled radial multiple-acquisition steady-state free precession for improved left atrial cine
Jaime S Robb1, Chenxi Hu2,3, Dana C Peters3
1University of Glasgow, Glasgow, UK.
Purpose:
Balanced steady-state free precession (bSSFP) left atrial (LA) cine suffers from off-resonance artifacts, particularly in the pulmonary veins (PVs). Linear combination or multiple-acquisition SSFP (MA-SSFP) effectively removes banding but greatly increases scan time. We hypothesized that MA-SSFP with interleaved radial undersampling, where each phase-cycling is acquired with an interleaved set of radial projections, would improve image quality of LA cine with a small increase of scan time and streak artefacts.
Methods:
Undersampled radial MA-SSFP with and without interleaving was compared with fully sampled radial bSSFP by means of simulations, phantoms, and in vivo imaging. Ten healthy subjects were imaged on a 3T scanner, with bSSFP and MA-SSFP cine of the left atrium, and B0-mapping. Images were assessed (1 = worst, 5 = best) by 2 independent readers, with respect to 5 qualitative criteria and apparent signal-to-noise ratio.
Results:
In healthy subjects, off-resonance differed from the right inferior PVs to the LA cavity by 163 Hz ± 73 Hz at 3T. Compared with fully sampled radial bSSFP, interleaved radial MA-SSFP significantly improved image quality with respect to off-resonance artifacts (3.8 ± 0.6 versus 2.3 ± 1.0; P = 0.005), PV conspicuity (2.8 ± 1.0 versus 4.3 ± 0.5; P = 0.005), and the number of visualized PVs (1.7 ± 0.4 versus 0.9 ± 0.7; P = 0.008), although with greater streak artifacts (3.4 ± 0.4 versus 4.9 ± 0.2; P = 0.004) and lower measured apparent signal-to-noise ratio (24 ± 9 versus 69 ± 36; P = 0.002). Flow artifacts were similar. Interleaved radial MA-SSFP reduced streaking artifacts and increased apparent signal-to-noise ratio versus noninterleaved radial.
Conclusions:
Interleaved radial MA-SSFP cine reduces banding artifacts with an acceptable increase of scan time and streak artifacts. The proposed technique improves the LA and PV visualization in bSSFP cine imaging.
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System V: CT