Method for a motion model based automated 4D dose calculation
Marc Ziegler1, Tobias Brandt1, Sebastian Lettmaier1
1Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Universitätsstraße 27, 91054 Erlangen, Germany.
Physics in Medicine and Biology
|October 17, 2019
Summary
Accurate 4D dose reconstruction for moving tumors using dynamic tumor tracking (DTT) requires a temporal resolution of 500 ms. This method ensures precise dose delivery in radiation therapy, improving treatment outcomes for patients with mobile tumors.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- The Vero system utilizes gimbaled dynamic tumor tracking (DTT) to treat moving tumors by adjusting the radiation beam.
- Tumor motion and changing beam geometry during DTT can affect the delivered radiation dose.
- Accurate dose reconstruction is crucial for evaluating and optimizing DTT treatments.
Purpose of the Study:
- To perform a comprehensive 4D dose reconstruction for patients treated with DTT.
- To determine the necessary temporal resolution for accurate 4D dose reconstruction.
- To assess the impact of DTT on dose distribution and identify factors influencing accuracy.
Main Methods:
- Developed a 4D motion model using 4D CT data to generate pseudo-CT images for various breathing phases.
- Calculated pseudo-CT images every 100 ms and performed dose calculations based on real-time beam geometry from log files.
- Accumulated dose distributions on the planning CT and compared parameters using gamma-evaluations at different spatial-temporal resolutions.
Main Results:
- Accumulated 4D dose distributions showed a mean Gross Tumor Volume (GTV) dose reduction of 0.77% compared to static plans.
- Observed larger dose deviations in some patients, particularly with lower 4D CT quality.
- Gamma-evaluation indicated that a temporal resolution of 500 ms is sufficient for accurate dose reconstruction.
Conclusions:
- A temporal resolution of 500 ms is adequate for precise 4D dose reconstruction in DTT treatments.
- Considering tumor motion with a spatial-temporal sampling of 1400 ms and 2 mm provides accurate results while reducing workload by 84%.
- Optimizing temporal and spatial resolution is key to balancing accuracy and efficiency in DTT-based radiotherapy.
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
230
Determining the optimal dose size and dosing frequency in pharmacotherapy is crucial for achieving therapeutic effectiveness while minimizing adverse effects. This article explores the methodologies employed in determining these parameters, focusing on their significance and interplay to tailor dosing regimens.Dose Size: Dose size refers to the amount of a drug administered in a single dose. It is determined based on the drug's pharmacodynamics and pharmacokinetics properties and...
230
Model Approaches for Pharmacokinetic Data: Distributed Parameter Models
216
Pharmacokinetic models are mathematical constructs that represent and predict the time course of drug concentrations in the body, providing meaningful pharmacokinetic parameters. These models are categorized into compartment, physiological, and distributed parameter models.
The distributed parameter models are specifically designed to account for variations and differences in some drug classes. This model is particularly useful for assessing regional concentrations of anticancer or...
The distributed parameter models are specifically designed to account for variations and differences in some drug classes. This model is particularly useful for assessing regional concentrations of anticancer or...
216