Twiddler's Syndrome: An Unusual Cause of Repeated Shocks by Implantable Cardioverter-Defibrillator in an Asymptomatic
Ahmad Jabri1, Zenab Laiq2, Yassar Nabeel3
1Department of Internal Medicine, Cleveland Clinic/Akron General Hospital, Akron, OH, USA.
Abstract:
Twiddler's syndrome refers to malfunction of pacemaker or implantable cardioverter-defibrillator (ICD) resulting from deliberate or unintentional twisting of pulse generator in device pocket, resulting in lead dislodgment. We report a case of a 65-year-old woman with multiple ICD shocks and antitachycardia pacing events on device interrogation, while she was unaware of the ICD discharges. Twiddler's syndrome is rare and mostly occurs in obese or elderly because of loose subcutaneous tissue allowing pulse generator to rotate in its pocket. It can be prevented by patient education against hand manipulating the generator. Smaller device pocket or suturing generator to fascia can also prevent future episodes.