Related Experiment Video
Updated: Jan 5, 2026

Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Use of MIRUS™ for MAC-driven application of isoflurane, sevoflurane, and desflurane in postoperative ICU patients: a
Martin Bellgardt1, Adrian Iustin Georgevici2, Mitja Klutzny2
1Department of Anaesthesiology and Intensive Care Medicine, Ruhr-University Bochum, St. Josef Hospital, Gudrunstraße 56, 44791, Bochum, Germany. m.bellgardt@klinikum-bochum.de.
Background:
The MIRUS™ (TIM, Koblenz, Germany) is an electronical gas delivery system, which offers an automated MAC (minimal alveolar concentration)-driven application of isoflurane, sevoflurane, or desflurane, and can be used for sedation in the intensive care unit. We investigated its consumption of volatile anesthetics at 0.5 MAC (primary endpoint) and the corresponding costs. Secondary endpoints were the technical feasibility to reach and control the MAC automatically, the depth of sedation at 0.5 MAC, and awakening times. Mechanically ventilated and sedated patients after major surgery were enrolled. Upon arrival in the intensive care unit, patients obtained intravenous propofol sedation for at least 1 h to collect ventilation and blood gas parameters, before they were switched to inhalational sedation using MIRUS™ with isoflurane, sevoflurane, or desflurane. After a minimum of 2 h, inhalational sedation was stopped, and awakening times were recorded. A multivariate electroencephalogram and the Richmond Agitation Sedation Scale (RASS) were used to assess the depth of sedation. Vital signs, ventilation parameters, gas consumption, MAC, and expiratory gas concentrations were continuously recorded.
Results:
Thirty patients obtained inhalational sedation for 18:08 [14:46-21:34] [median 1st-3rd quartiles] hours. The MAC was 0.58 [0.50-0.64], resulting in a Narcotrend Index of 37.1 [30.9-42.4] and a RASS of - 3.0 [- 4.0 to (- 3.0)]. The median gas consumption was significantly lowest for isoflurane ([ml h-1]: isoflurane: 3.97 [3.61-5.70]; sevoflurane: 8.91 [6.32-13.76]; and desflurane: 25.88 [20.38-30.82]; p < 0.001). This corresponds to average costs of 0.39 € h-1 for isoflurane, 2.14 € h-1 for sevoflurane, and 7.54 € h-1 for desflurane. Awakening times (eye opening [min]: isoflurane: 9:48 [4:15-20:18]; sevoflurane: 3:45 [0:30-6:30]; desflurane: 2:00 [1:00-6:30]; p = 0.043) and time to extubation ([min]: isoflurane: 10:10 [8:00-20:30]; sevoflurane: 7:30 [4:37-14:22]; desflurane: 3:00 [3:00-6:00]; p = 0.007) were significantly shortest for desflurane.
Conclusions:
A target-controlled, MAC-driven automated application of volatile anesthetics is technically feasible and enables an adequate depth of sedation. Gas consumption was highest for desflurane, which is also the most expensive volatile anesthetic. Although awakening times were shortest, the actual time saving of a few minutes might be negligible for most patients in the intensive care unit. Thus, using desflurane seems not rational from an economic perspective. Trial registration Clinical Trials Registry (ref.: NCT03860129). Registered 24 September 2018-Retrospectively registered.
More Related Videos
03:02Application of Dixon's Up-and-Down Design to Estimate the Minimum Alveolar Concentration of Sevoflurane in Rats with Refined Movement Classification
Published on: July 25, 2025
04:56In Vitro Method to Control Concentrations of Halogenated Gases in Cultured Alveolar Epithelial Cells
Published on: October 23, 2018
Related Concept Videos
Inhalational Anesthetics: Overview
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...