Feasibility of two-dimensional dose distribution deconvolution using convolution neural networks.
Wonjoong Cheon1, Sung Jin Kim2, Kyuseok Kim3
1Department of Health Sciences and Technology, SAIHST, Sungkyunkwan University, Seoul, 06351, Korea.
Convolutional neural networks (CNNs) show promise for improving two-dimensional (2D) dose distribution deconvolution in radiation therapy. The developed PenumbraNet significantly enhances accuracy compared to traditional analytical methods.
Area of Science:
- Medical Physics
- Radiotherapy
- Machine Learning in Healthcare
Background:
- Accurate dose distribution is critical for effective radiotherapy.
- Scintillation detectors can introduce artifacts, particularly in the penumbra region.
- Traditional analytical methods for dose deconvolution have limitations.
Purpose of the Study:
- To evaluate the feasibility of using convolutional neural networks (CNNs) for two-dimensional (2D) dose distribution deconvolution.
- To address detector-interface artifacts in the penumbra region of an in-house scintillation detector.
- To compare CNN-based deconvolution with analytical approaches.
Main Methods:
- Developed a shallow linear CNN, PenumbraNet, for dose deconvolution.
- Trained PenumbraNet on datasets from a Novalis Tx medical linear accelerator, including square fields and IMRT plans.
- Validated PenumbraNet using test data and evaluated performance with gamma index criteria (2%/2mm, 3%/3mm), comparing with analytical Wiener filtering and different CNN architectures.
Main Results:
- PenumbraNet achieved a mean gamma passing rate of 95.81% (3%/3mm), significantly outperforming analytical Wiener filtering (84.77%).
- Nonlinear PenumbraNet variants showed even higher performance, with the best achieving 96.62% (3%/3mm).
- The best performing nonlinear PenumbraNet improved the gamma passing rate by 11.85% compared to the measured dose distribution.
Conclusions:
- Demonstrated the feasibility and effectiveness of PenumbraNets for 2D dose distribution deconvolution.
- CNN-based deconvolution, particularly nonlinear variants, offers superior accuracy in correcting for detector artifacts.
- This approach holds potential for enhancing dose calculation accuracy in clinical radiotherapy.
More Related Videos
07:013D Imaging of Soft-Tissue Samples using an X-ray Specific Staining Method and Nanoscopic Computed Tomography
Published on: October 24, 2019
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
Related Concept Videos
Deconvolution
Deconvolution involves several mathematical techniques to derive the impulse response. One common approach is polynomial division. In this method, the input and output sequences are treated as coefficients of...
Dose Size and Dosing Frequency: Determination Methods