Related Experiment Video
Updated: Jan 5, 2026

Metacarpal Small Incision for Carpal Tunnel Syndrome
Published on: April 5, 2024
Median Nerve Injury Caused by Screw Malpositioning in Percutaneous Scaphoid Fracture Fixation
Marta Starnoni1, Giulia Colzani2, Giorgio De Santis1
1Division of Plastic Surgery, University of Modena and Reggio Emilia, Modena, Italy.
Abstract:
An unusual complication of percutaneous fixation of a schapoid fracture in a 36-year-old man is described. After this surgery, the patient complained of numbness and tingling within the median nerve distribution and thenar weakness. An Electromyography showing moderate CTS had led to carpal tunnel decompression, which resulted unsuccessful. When the patient presented to us, both sensory and motor median nerve impairment were present, interfering with his professional activity. A computed tomography scan performed to assess the scaphoid screw position, showed that the screw had been placed volar to the scaphoid, adjacent to the median nerve. Surgery confirmed median nerve compression by the screw, especially during complete wrist extension. The surgical removal of the screw led to immediate improvement of both the motor and sensory functions of the nerve, with complete clinical restoration of nerve function and resolution of the symptoms 2 months postoperatively.