Related Experiment Video
Updated: Jan 5, 2026

Using Enzyme-based Biosensors to Measure Tonic and Phasic Glutamate in Alzheimer's Mouse Models
Published on: May 3, 2017
Using Amperometric, Enzyme-Based Biosensors for Performing Longitudinal Measurements of Extracellular Adenosine
1Department of Physiology and Medical Physics, Royal College of Surgeons in Ireland, Dublin 2, Ireland. edwardbeamer@rcsi.com.
Abstract:
Adenosine 5-triphosphate (ATP) functions in the central nervous system as an extracellular signaling molecule. While much progress has been made in understanding the circumstances under which it is released, from in vitro preparations, in vivo has proven more challenging. Microdialysis followed by high-performance liquid chromatography has been employed to demonstrate a spike in extracellular concentrations under some pathological conditions including seizures, but this method lacks the sensitivity to detect extracellular ATP at concentrations present under normal physiological conditions. An alternative approach, the use of amperometric, enzyme-based microelectrode biosensors for measuring extracellular ATP in vivo have been employed in the rabbit. Here, we describe a protocol for measuring ATP concentrations using these biosensors in the mouse, simultaneously with electroencephalogram recordings. This approach is ideal for investigating the relationship between ATP release and seizures.