SVT discrimination algorithms significantly reduce the rate of inappropriate therapy in the setting of modern-day
Alan Cheng1, Angelo Auricchio2, Edward J Schloss3
1Medtronic, plc, Mounds View, Minnesota.
Background:
Contemporary implantable cardioverter-defibrillator (ICD) programming involving delayed high-rate detection and use of supraventricular tachycardia (SVT) discriminators has significantly reduced the rate of inappropriate shocks. The extent to which SVT algorithms alone reduce inappropriate therapies is poorly understood.
Methods And Results:
PainFree SST enrolled 2770 patients with a single- or dual-chamber ICD or cardiac resynchronization defibrillator. Patients were followed for 22 ± 9 months with SVT discriminators on in 96% of patients. Sustained ventricular tachyarrhythmias and SVT episodes were adjudicated by an independent physician committee. For this analysis, all episodes were subjected to postprocessing computer simulation with SVT discriminators off with and without delayed high-rate detection criteria (ventricular fibrillation zone only, 30/40 at 320 ms). There were 3282 adjudicated SVT episodes of which 115 resulted in an ICD shock and 113 received only ATP (2-year inappropriate shock and therapy rates of 3.1% and 4.1%). Therapy was appropriately withheld for the remaining 3054 SVT episodes. With both SVT discriminators and delayed high-rate detection simulated off, the 2-year inappropriate therapy rate would have been 22.9% (hazard ratio [HR] = 6.24; 95% confidence interval [CI]: 5.20-7.49). With SVT discriminators simulated off and delayed high-rate detection simulated on in all patients, the 2-year rate would have been 6.4% (HR = 1.63; CI: 1.44-1.85).
Conclusions:
The use of SVT discriminators has a significant role in reducing the rate of inappropriate ICD therapy even in the setting of delayed high-rate detection settings. Deactivating SVT discriminators would have resulted in an overall increase in the inappropriate ICD therapy rate by 63% and 524% with and without delayed high-rate detection programming, respectively.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
10:50Behavioral Determination of Stimulus Pair Discrimination of Auditory Acoustic and Electrical Stimuli Using a Classical Conditioning and Heart-rate Approach
Published on: June 6, 2012
Related Concept Videos
Dysrhythmias VI: Management of Dysrhythmias
Antiarrhythmic Drugs: Class IV Agents as Calcium Channel Blockers
Verapamil, a calcium channel blocker, inhibits calcium movement across myocardial cell membranes and vascular smooth muscle. This results in the dilation of coronary and...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiomyopathy III: Hypertrophic Cardiomyopathy