Comparison of image quality between split-filter twin beam dual energy and single energy images in abdominal CT
Zhongfeng Niu1, Jiao Chen1, Hong Ren1
1Department of Radiology, Sir Run Run Shaw Hospital, Zhejiang University College of Medicine and Sir Run Run Shaw Institute of Clinical Medicine of Zhejiang University, No. 3, East Qingchun Road, Hangzhou, 310016, China.
Purpose:
To compare the objective and subjective image quality between composed images from split-filter twin beam dual energy (TBDE) and single-energy computed tomography (SECT) in abdominal CT.
Methods:
In this prospective study, 103 patients were imaged using TBDE (n = 51) or SECT (n = 52). The CT number and noise were measured for the following six abdominal structures: liver, spleen, fat, muscle, aorta and portal vein. The normalised noise level for the liver was separately measured and compared. The consistency of the SNR and CT number was compared between the two groups. The subjective image quality was evaluated using six aspects in a blinded manner. Cohen's Kappa statistic was used to determine the level of agreement between the two radiologists.
Results:
For the objective image quality comparison, the SNR of all structures was higher using TBDE compared to SECT (p < 0.05). The CT value for different structures were comparable between the two groups (p > 0.05). Among all patient sizes, the noise level for TBDE images was significantly lower (7-17% reduction) compared to the SECT images (p < 0.01). Furthermore, noise reduction's magnitude increases with body size. For image quality's subjective evaluation, TBDE images are superior for certain aspects. Cohen's Kappa values (0.7634-0.8460) suggest an adequate level of agreement between the two observers.
Conclusions:
TBDE scan mode can yield similar or even better objective and subjective image quality at the same level of radiation than conventional SECT. Quantitatively, TBDE images have a 7-17% reduction in noise, depending on the size of the scanned body regions.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
X-ray Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...