Purpose:
To determine liquid nitrogen evaporation rates of intact liquid nitrogen storage tanks and tanks with their vacuum removed.
Methods:
Donated storage tank performance (LN2 evaporation) was evaluated before and after induced vacuum failure. Vacuum of each tank was removed by drilling through the vacuum port. Temperature probes were placed 2 in. below the bottom of the styrofoam cap/plug, and tanks were weighed every 3 h. Evaporation rate and time from failure to the critical temperature was determined.
Result:
Storage tanks with failed vacuum have a much higher evaporation rate than those with intact vacuum; evaporation rates increased dramatically within 3 to 6 h in the smaller tanks, and time to complete depletion varied according to starting LN2 volume. Tanks with storage racks/specimens may have altered evaporation profiles compared to tanks without. Locating temperature probes 2 in. below the styrofoam cap/plug suggests that for most applications, alarms would sound approximately 1 h prior to reaching the critical warming temperature, approximately - 130 °C. External signs of vacuum loss were dramatic: vapor, frost, and audible movement of air.
Conclusion:
For the first time, we have data on how liquid nitrogen storage tanks behave when their vacuum is removed. These findings are conservative; each lab must consider starting volume, tank size/capacity, function (storage or shipping), age, and pre-existing evaporation behavior in order to develop an emergency response to critical tank failure. Times to complete failure/evaporation and critical warming temperature after vacuum loss are different; these data should be considered when evaluating tank alarm systems.
Related Concept Videos
Thermal Expansion
Frost Action on Concrete
This freeze-thaw cycle primarily causes surface scaling, where...
Constant Volume Calorimetry
Cold Weather Concreting
To counteract the negative impacts of cold weather, ensuring...
Thermal expansion and Thermal stress: Problem Solving
To solve the problem, first, identify the known and unknown quantities. The initial length (L) of the bridge is 1275 m, the coefficient of linear expansion (α) for steel is 12 x 10-6/°C, and the change in temperature (ΔT) is 55...