Effect of iterative reconstruction techniques on image quality in low radiation dose chest CT: a phantom study
Yan Xu1, Ting-Ting Zhang1, Zhi-Hai Hu1
1Department of Radiology, Beijing Friendship Hospital, Capital Medical University, Beijing, China.
Purpose:
We aimed to evaluate the quality of chest computed tomography (CT) images obtained with low-dose CT using three iterative reconstruction (IR) algorithms.
Methods:
Two 64-detector spiral CT scanners (HDCT and iCT) were used to scan a chest phantom containing 6 ground-glass nodules (GGNs) at 11 radiation dose levels. CT images were reconstructed by filtered back projection or three IR algorithms. Reconstructed images were analyzed for CT values, average noise, contrast-to-noise ratio (CNR) values, subjective image noise, and diagnostic acceptability of the GGNs. Repeated-measures analysis of variance was used for statistical analyses.
Results:
Average noise decreased and CNR increased with increasing radiation dose when the same reconstruction algorithm was applied. Average image noise was significantly lower when reconstructed with MBIR than with iDOSE4 at the same low radiation doses. The two radiologists showed good interobserver consistency in image quality with kappa 0.83. A significant relationship was found between image noise and diagnostic acceptability of the GGNs.
Conclusion:
Three IR algorithms are able to reduce the image noise and improve the image quality of low-dose CT. In the same radiation dose, the low-dose CT image quality reconstructed with MBIR algorithms is better than that of other IR algorithms.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography