Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic
Seong-Hi Park1, Young-Su Park2
1School of Nursing, Soonchunhyang University, Asan.
Objective:
A mercury sphygmomanometer has been considered a gold standard for measuring blood pressure. However, by the Minamata Convention on Mercury, the traditional mercury sphygmomanometer is being replaced by an automated oscillometric device. This study aimed to provide scientific evidence to determine whether an automated oscillometric device can replace a mercury sphygmomanometer and if it is applicable in routine practice.
Methods:
MEDLINE, EMBASE, the Cochrane Library, and CINAHL were searched on 4 May 2018. Studies comparing blood pressure measurements between automated oscillometric devices and mercury sphygmomanometers were included. Study characteristics were abstracted using the evidence table, and random-effects meta-analyses were conducted.
Results:
Data were compiled from 24 studies comprising 47 759 subjects. The results of meta-analysis showed that automated oscillometric devices measured lower than mercury sphygmomanometers for both systolic blood pressure (mean differences -1.75 mmHg, 95% confidence intervals: -3.05 to -0.45, I = 91.0%) and diastolic blood pressure (mean differences -1.20 mmHg, 95% confidence intervals: -2.16 to -0.24, I = 95.0%). In sub-group analyses by manufacturer, BpTRU measured lower than the mercury sphygmomanometer and OMRON showed no difference compared to the mercury sphygmomanometer for both systolic and diastolic blood pressure, but the results differed depending on the devices.
Conclusion:
As a result of this review, the difference in blood pressure between the mercury sphygmomanometer and the automated oscillometric device was within 5 mmHg, but the heterogeneity between the studies was very high. The automated oscillometric devices showed differences in blood pressure results according to the manufacturer and product type.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...