Ion-Neutral Collision Effects on Ion Trapping and Pseudopotential Depth in Ion Trap Mass Spectrometry
Ming Li1,2, Xinwei Liu2, Xiaoyu Zhou3
1NCS Testing Technology CO., Ltd, Beijing, 100081, China.
Abstract:
Ion trapping using radio-frequency (RF) devices has been widely used in mass spectrometry (MS). The pseudopotential well (PW) model enables the use of a pseudopotential depth, D, to evaluate the ion trapping capability of the RF devices in the pure electric field. It remains unclear how gas pressures regulate the ion trapping and D. Here, we calculated the D of a linear ion trap (LIT) from 1 mTorr to 2 Torr, a pressure range critical for the operation of the RF devices, through ion cloud simulations. Compared with the case of pure electric field, ion-neutral collision effects at pressures of 1 to 100 mTorr were beneficial for the ion trapping and revealed an optimal trapping depth, D, at around 10 mTorr. We explained the mechanism and validated the observation via ion trapping experiments performed in a home-made dual LIT mass spectrometer. We also showed that near the stability boundary, the RF heating became comparable with the D, which led to the decrement of ion trapping capability characterized by the available D.
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometers
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Chemical Ionization (CI) Mass Spectrometry