Related Experiment Video
Updated: Jan 4, 2026

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Effective Dose Assessment for Patients Undergoing Contemporary Fluoroscopically Guided Interventional Procedures
Xinhua Li1, Joshua A Hirsch1, Madan M Rehani1
1Department of Radiology, Massachusetts General Hospital, 55 Fruit St, Boston, MA 02114.
This study establishes air kerma-area product (KAP) and effective dose benchmarks for 101 fluoroscopically guided interventional procedures. These values aid in clinical research and dose management where national benchmarks are lacking.
Area of Science:
- Medical Physics
- Radiology
- Radiation Safety
Background:
- Fluoroscopically guided interventional (FGI) procedures are increasingly common, necessitating accurate radiation dose assessment.
- Establishing procedure-specific radiation dose metrics is crucial for patient safety and quality improvement.
- Existing national benchmarks for air kerma-area product (KAP) and effective dose are limited for many FGI procedures.
Purpose of the Study:
- To determine procedure-specific air kerma-area product (KAP) and effective dose for a wide range of FGI procedures.
- To provide data that can be used for clinical research protocol design and dose management programs.
- To assist healthcare organizations in assessing compliance with radiation safety standards.
Main Methods:
- Retrospective analysis of dose data from 24,911 adult FGI cases across 12 examination rooms.
- Calculation of effective dose using KAP and conversion coefficients, primarily from NCRP Report 160.
- Statistical analysis to determine mean values and percentiles (10th, 25th, 50th, 75th, 95th) for KAP and effective dose.
Main Results:
- Procedure-specific KAP and effective dose values were established for 101 FGI procedures.
- National benchmarks were unavailable for 89 KAP values and all 101 effective dose values.
- Twelve common procedures showed median KAP < 3.26 Gy·cm² and median effective dose < 0.70 mSv, while some infrequent procedures had higher doses (e.g., portography 95th percentile KAP ≥ 985 Gy·cm², effective dose ≥ 256 mSv).
Conclusions:
- The study provides essential procedure-specific radiation dose data (KAP and effective dose) for numerous FGI procedures.
- These data are valuable for designing clinical research, implementing dose management, and meeting regulatory compliance in the absence of national benchmarks.
- The findings highlight the variability in radiation doses among FGI procedures, underscoring the need for tailored dose monitoring and optimization strategies.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies VII: Vascular Imaging
Imaging Studies III: Computed Tomography
Cardiac Catheterization I: Pre-Procedure Overview