Standing Neurophysiological Assessment of Lower Extremity Muscles Post-Stroke
Published on: July 26, 2021
Surface electromyography of the foot: A protocol for sensor placement
Helen Branthwaite1, Christopher Aitkins1, Steven Lindley2
1Centre for Biomechanics and Rehabilitation Technologies, Staffordshire University, Stoke on Trent, UK.
Surface EMG (sEMG) shows inconsistent reliability for assessing intrinsic foot muscles. Further research is needed to refine protocols for accurate muscle activity measurement.
Area of Science:
- Biomechanics
- Neuroscience
- Musculoskeletal research
Background:
- Surface electromyography (sEMG) is widely used for muscle activity recording.
- Previous studies on intrinsic foot muscles using sEMG were limited due to technological constraints, often requiring fine-wire electrodes.
- This study addresses the need for reliable sEMG protocols for intrinsic foot muscles.
Purpose of the Study:
- To evaluate the reliability of a surface EMG protocol for assessing intrinsic foot muscles.
- To establish a standardized sEMG protocol for intrinsic foot muscle analysis.
- To identify potential challenges and areas for improvement in sEMG application for foot muscles.
Main Methods:
- Six intrinsic foot muscles were identified using ultrasound and functional testing.
- A standardized sEMG sensor placement protocol was developed.
- Repeatability and reliability were assessed across multiple testers, subjects, and time sessions using Intraclass Correlation Coefficient (ICC) and percentage standard error of measurement.
Main Results:
- Good correlation was observed for Extensor Digitorum Brevis, Dorsal Interossei, Abductor Digiti Minimi, and Flexor Digitorum Brevis.
- Increased variability and poor correlation were found for Flexor Hallucis Brevis and Abductor Hallucis.
- Low precision of measurement was indicated by the percentage standard error of measurement, despite high ICC values.
Conclusions:
- Inconsistent reliability of sEMG for intrinsic foot muscles was demonstrated due to variability between testers and sessions.
- Further research and protocol refinement, potentially focusing on muscle grouping, are necessary to improve understanding and measurement accuracy.
- The findings highlight the need for optimized sEMG techniques for reliable intrinsic foot muscle assessment.