Purpose:
Proximal ulna fractures are common injuries and frequently treated with angular stable plating. This surgical option shows good functional results. Relevant drawbacks such as large soft tissue exposure, compromised blood supply of fracture fragments and disturbing osteosynthetic material are described. The aim of this study was to compare a new locked proximal ulna nail with angular stable plating in a biomechanical testing setup for extraarticular proximal ulna fractures.
Methods:
Ten pairs of sawbones with a Jupiter type IIB proximal ulna fracture (OTA 2U1A3.1) were tested after osteosynthesis with the mentioned implants in a servo-pneumatic testing machine. The testing setup simulates physiological joint motion (0°-90°) under cyclic loading (30-300 N). Primary stability and loosening of both constructs were quantified using micromotion video-analysis after 608 test cycles.
Results:
The micromotion analysis showed significantly higher primary stability of the locked nail (0.29 ± 0.13 mm) compared to the angular stable plate (0.97 ± 0.30 mm, p < 0.001). Both implants showed a low amount of loosening after completion of the test cycles. The construct with the locked nail (0.08 ± 0.06 mm) showed significantly lower dislocation of the fragments measured at the anterior cortex (plate 0.24 ± 0.13 mm, p < 0.001).
Conclusion:
Nailing of proximal ulna fractures shows significantly higher primary stability and lower loosening compared to angular stable plating in our testing setup.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Bones of the Upper Limb: Ulna
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...