Vestibular Implantation and the Feasibility of Fluoroscopy-Guided Electrode Insertion
Joost Johannes Antonius Stultiens1, Alida Annechien Postma2, Nils Guinand3
1Department of Otorhinolaryngology-Head and Neck Surgery, School for Mental Health and Neuroscience, Faculty of Health Medicine and Life Sciences, P.O. box 5800, 6202 AZ, Maastricht University Medical Center, Maastricht, The Netherlands.
Abstract:
Recent research has shown promising results for the development of a clinically feasible vestibular implant in the near future. However, correct electrode placement remains a challenge. It was shown that fluoroscopy was able to visualize the semicircular canal ampullae and electrodes, and guide electrode insertion in real time. Ninety-four percent of the 18 electrodes were implanted correctly (<1.5 mm distance to target). The median distances were 0.60 mm, 0.85 mm, and 0.65 mm for the superior, lateral, and posterior semicircular canal, respectively. These findings suggest that fluoroscopy can significantly improve electrode placement during vestibular implantation.