First Trimester Mean Arterial Pressure Measured Manually Versus Using an Automated Device and the Prediction of
Maxime Côté1, Amélie Boutin2, Cédric Gasse3
1Department of Gynecology, Obstetrics and Reproduction, Faculty of Medicine, Université Laval, Québec City, QC.
Objective:
First trimester mean arterial blood pressure (MAP) can be used to predict preeclampsia. This study aimed to compare the performance of first trimester MAP measured with an automated device using a standardized technique versus MAP taken manually in a typical clinical setting.
Methods:
A case-cohort study niched into a prospective cohort of pregnant women recruited at 11-14 weeks was performed. MAP was measured with an automated device on both arms until stability was reached. These results were compared with the MAP measured with a manual device at the closest medical visit (between 10 and 15 weeks gestation) and noted in the medical charts. Receiver-operator characteristics curve analyses were used to estimate the predictive values of MAP measured by both techniques.
Results:
Forty-one women with preeclampsia and 167 control patients were used for the comparisons. MAP measured with an automated device decreased significantly between 11 and 14 weeks gestation (P < 0.001). Moreover, MAP measured with an automated device was a better predictor of preeclampsia (area under the curve 0.70; 95% confidence interval 0.61-0.79) than MAP measured with a manual device in a clinical setting (area under the curve 0.60; 95% confidence interval 0.50-0.70). Taken alone, MAP measured with an automated device was associated with a detection rate of preeclampsia of 34%, for a false-positive rate of 10%.
Conclusion:
First trimester MAP can predict preeclampsia. This study demonstrated that MAP measured with an automated device using a standardized technique is a better predictor than MAP measured with a manual device.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: