Effective Radiation Dose Reduction in Computed Tomography With Iterative Reconstruction in Patients With Urinary
1From the Department of Radiology, School of Medicine, Kyungbook National University, Kyungbook National University Hospital, Jung-gu, Daegu, Korea.
Purpose:
This study aimed to prospectively compare the image quality and visibility of urinary stone on computed tomographic (CT) images at multiple radiation exposure levels from the same patient reconstructed with sinogram-affirmed iterative reconstruction (SAFIRE).
Methods:
This study included 760 patients with urinary stone who underwent CT with simultaneous acquisition at 6 exposures per patient (100% filtered back projection, 75%, 50%, 37.5%, 25%, and 12.5% SAFIRE). Two radiologists independently assessed overall image quality, noise, and stone visibility by using a 5-point scale. Quantitative measurements, including the CT number, image noise, signal-to-noise ratio, contrast-to-noise ratio (CNR), and corresponding figure of merit (FOM), were compared for 100% versus 5 other radiation doses.
Results:
Qualitative overall image quality, noise, and stone visibility according to the location were not inferior at 37.5% exposure compared with 100% exposure, except for the visualization of smaller stones <3 mm. The signal-to-noise ratio and CNR of CT images were increased at 50% exposure compared with 100% exposure. Computed tomographic images at 37.5% exposure reconstructed with SAFIRE had significantly more noise and a lower CNR compared with CT images reconstructed with filtered back projection, based on FOMnoise and FOMCNR. The size-specific dose estimation was 4.1 ± 0.8 mGy at 37.5% exposure.
Conclusion:
Computed tomography performed at 37.5% exposure with SAFIRE may be diagnostically acceptable for the detection of clinically relevant stone.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Urinary Tract Calculi III: Medical Management
Imaging Studies IV: Magnetic Resonance Imaging
Urinary Tract Calculi IV: Nutrition Therapy and Prevention
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...