Related Experiment Video
Updated: Jan 4, 2026

A Method to Test the Efficacy of Handwashing for the Removal of Emerging Infectious Pathogens
Published on: June 7, 2017
Comparative assessment of the effectiveness of three disinfection protocols for reducing bacterial contamination of
Patricia Sebastian Marcos1, Darren Hermes2, Mellora Sharman1
1Small Animal Referral Centre, Animal Health Trust, Kentford, Suffolk, United Kingdom.
Abstract:
In a crossover study, 30 stethoscopes were assessed and disinfected using 3 protocols: isopropyl alcohol, a quaternary ammonia or biguanide disinfectant, and ultraviolet germicidal irradiation (UVGI). All protocols effectively reduced bacterial loads, but UVGI was less effective at higher contamination levels (P = .0004). The effectiveness of each intervention was short in duration.
Insights
Disinfection protocols for stethoscopes, including alcohol, disinfectants, and UV germicidal irradiation (UVGI), effectively reduced bacteria. However, UVGI showed reduced efficacy with higher contamination, and all methods provided only short-term effectiveness.
Area of Science:
- Infection Control
- Microbiology
- Medical Device Disinfection
Background:
- Stethoscopes are critical medical tools frequently contaminated with bacteria.
- Effective disinfection of stethoscopes is essential to prevent healthcare-associated infections.
- Assessing various disinfection methods is crucial for optimizing patient safety.
Purpose of the Study:
- To compare the efficacy of three common stethoscope disinfection protocols.
- To evaluate the effectiveness of isopropyl alcohol, quaternary ammonia/biguanide disinfectants, and UV germicidal irradiation (UVGI).
- To determine the duration of bacterial load reduction for each method.
Main Methods:
- A crossover study design was employed.
- Thirty stethoscopes were subjected to disinfection using three distinct protocols.
- Bacterial loads were assessed before and after disinfection to measure efficacy.
Main Results:
- All tested disinfection protocols demonstrated significant bacterial load reduction.
- UV germicidal irradiation (UVGI) was less effective in reducing bacterial loads at higher contamination levels (P = .0004).
- The antimicrobial effectiveness of all interventions was transient, indicating short-term efficacy.
Conclusions:
- Isopropyl alcohol, chemical disinfectants, and UVGI are effective in reducing stethoscope bacterial contamination.
- UVGI's effectiveness is compromised by high bacterial loads.
- The short duration of effectiveness for all methods highlights the need for frequent disinfection protocols.
More Related Videos
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Antimicrobial Effectiveness
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Hand hygiene
Hand washing...