Evaluation of Visual Serum Indices Measurements and Potential False Result Risks in Routine Clinical Chemistry Tests
Tigist Getahun1, Anberber Alemu1, Firehiwot Mulugeta1
1Ethiopian Public Health Institute, National References Laboratory for Clinical Chemistry, Addis Ababa, Ethiopia.
Background:
Serum indices (SI) including hemolyzed, lipemic, and icteric samples, affects the accuracy of test result. The aim of this study was to evaluate SI values done by visual inspections and potential false result risks by comparing with actual measurements done by Cobas 6000 Chemistry analyzer at Ethiopian Public Health Institute (EPHI), Addis Ababa, Ethiopia.
Methods:
An observational, cross-sectional study was conducted from April to May 2017 on samples referred to Clinical Chemistry laboratory of EPHI, Ethiopia. These samples SI values, after visual inspection by three trained observers, was analyzed again on Roche Cobas 6000 analyzer (RCA). The generated data was analyzed by using weighted kappa methods on STAT statistical software version 20.
Results:
From a total of 1509 samples, SI values identified by the RCA as hemolysis, icteric, and lipemic were 933 (62%), 74(5%) and 59(4%) respectively. The SI average weighted kappa between RCA and visual inspection were: 0.1870, 0.3421, and 0.1259 for hemolysis, icteric, and lipemic samples, respectively. Combined inter-observers variability among observers for hemolysis, Icterus, and lipemic samples were 0.4758, 0.3258, and 0.3628 respectively. The best agreement among observers was in the case of hemolysis (0 grades), while the lowest agreement was observed in the case of icterus (+3 grades). In addition, test parameters, such as CK-MB (22%), and LDH (20%) were falsely accepted, whereas Cl- and Na+ (up to 25%) were falsely rejected tests by observers. On the other hand, results rejected by Cobas SI assessments included CK-MB (22%), LDH (20%), and BIL-D (4%).
Conclusion:
Visual inspection of SI showed poorly agreement with automated system. Thus, there is genuine need for more training of Laboratory professionals on identification of SI, and as much as possible SI should be done by automated system to improve quality of test results.
Related Concept Videos
Serum Studies: Renal Function Tests
Serum Laboratory Studies, Stool Test, Breath Test
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...