Effect of blood viscosity on the performance of virtually wall-less venous cannulas
Saad Abdel-Sayed1, Philippe Abdel-Sayed2, Denis Berdaj3
1Department of Surgery and Anesthesiology, CHUV, Lausanne, Switzerland.
Aim:
This study was designed to quantify the influence of blood as test medium compared to water in cannula bench performance assessment.
Methods:
An in vitro circuit was set-up with silicone tubing between two reservoirs. The test medium was pumped from the lower reservoir by centrifugal pump to the upper reservoir. The test-cannula was inserted in a silicone tube connected between the lower reservoir and the centrifugal pump. Flow rate and pump inlet-pressure were measured for wall-less versus thin-wall cannula using a centrifugal pump in a dynamic bench-test for an afterload of 40-60 mmHg using two media: blood 10 g/dL and 5.6 g/dL and water 0 g/dL.
Results:
The wall-less cannula showed significantly higher flows rates as compared to the thin-wall cannula (control), with both hemoglobin concentrations and water. Indeed, for a target volume of 200-250 mL of blood (Hg 10 g/dL) in the upper reservoir, the cannula outlet pressure (P) was -14 ± 14 mmHg versus -18 ± 11 mmHg for the wall-less and control respectively; the cannula outlet flow rate (Q) was 3.91 ± 0.41 versus 3.67 ± 0.45 L/min, respectively. At the same target volume but with a Hg of 5.7 g/dL, P was -16 ± 12 mmHg versus -19 ± 12 mmHg and Q was 4 ± 0.1 versus 4 ± 0.4 L/min for the wall-less cannula and control respectively. Likewise, P and Q values with water were -1 mmHg versus -0.67 ± 0.58 mmHg and 4.17 ± 0.45 L/min versus 4.08 ± 0.47 L/min for the wall-less and control respectively.
Conclusion:
Walls-less cannula showed 5.6% less pump inlet-pressure differences calculated between blood and water, as compared to that of thin-wall cannula (-21 times). Flow differences were 6% and 10% for the walls-less and thin-wall cannula respectively. We conclude that testing the cannula performance with water is a good scenario and can overestimate the flow by a 10%. However, superiority for wall-less is preserved with both water and blood.
More Related Videos
Related Concept Videos
Veins as Blood Reservoirs
Viscosity
The SI unit of viscosity is...
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Venules
Venules are much smaller in diameter compared to their larger counterparts, the veins. They are generally 8 to 100 micrometers in diameter, significantly smaller than the size of veins. The walls of venules are thin, consisting of the endothelium, a thin layer of connective tissue, and occasionally a few smooth muscle cells. This structural simplicity is a stark contrast...
Vascular Resistance
The primary determinants of vascular resistance are vessel diameter, blood viscosity, and vessel length. Among these, vessel diameter plays the most significant role due to the fourth power relationship described by...