ANGULAR DEPENDENCE OF TRACK-ETCH DETECTOR HARZLAS TD-1
Iva Ambrožová1, Kateřina Pachnerová Brabcová1, Vyacheslav A Shurshakov2
1Department of Radiation Dosimetry, Nuclear Physics Institute of the Czech Academy of Sciences, Na Truhlářce 39/64, 18000 Prague, Czech Republic.
Abstract:
Track-etched detectors are commonly used also for radiation monitoring onboard International Space Station. To be registered in track-etched detectors, the particle needs to meet several criteria-it must have linear energy transfer above the detection threshold and strike the detector's surface under an angle higher than the so-called critical angle. Linear energy transfer is then estimated from calibration curve from the etch rate ratio V that is calculated from parameters of individual tracks appearing on the detector's surface after chemical etching. It has been observed that V can depend on the incident angle and this dependence can vary for different detector materials, etching and evaluating conditions. To investigate angular dependence, detectors (Harzlas TD-1) were irradiated at HIMAC by several ions under angles from 0° to 90°. The correction accounting not only for critical angle but also for dependence of V on the incident angle is introduced and applied to spectra measured onboard International Space Station.