Relative response of dosimeters to variations in scattered X-ray energy spectra encountered in interventional
Maeve Masterson1, Seán Cournane2, Nina McWilliams2
1School of Physics, National University of Ireland, Galway, Ireland.
Purpose:
The new lower eye lens dose limit is of relevance in interventional radiology, where higher dose procedures result in increased scattered radiation to staff. The eye lens dose may be monitored using the directional dose equivalent at 3 mm depth, Hp(3), or through Hp(10) or Hp(0.07) measurements and using conversion factors. However, there are a considerable range of factors which contribute to measurement uncertainties, one of which is the incident photon energy. This study investigated the energy spectra of scattered radiation in interventional radiology, and the dosimetry accuracy of dosimeter types, evaluating their energy dependence.
Methods:
Scatter X-ray energy spectra were recorded under varied conditions in a fluoroscopy imaging suite. Dosimetry accuracy of eye dosimeters, including TLDs (100 s, 100Hs), Landauer Hp(3), John Caunt ED3 and Electronic Personal dosimeters (EPDs) were compared to air kerma measurements across a range of tube voltages.
Results:
The variation of energy spectra with changing phantom thickness, spectrometer angulation and filtration are presented. The 100 and 100H TLDs, and EPDs showed a consistent air kerma response (within 10%) with changes in energy. The real-time silicon diode detectors showed a variable over response of between 10 and 25% across the energies investigated while Landauers dedicated Hp(3) eye dosimeters showed considerable variation between dosimeters for similar conditions, a 17% variation at 50 kVp.
Conclusion:
The work aimed to validate the scattered energy spectra typically encountered in interventional radiology and to further determine the accuracy of eye dosimeters in relation to energy response variations.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Biological Effects of Radiation
X-ray Imaging
Dose-Response Relationship: Overview
Dosage Regimen: Individualization
Dose-Response Relationship: Selectivity and Specificity
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...