Related Experiment Video
Updated: May 7, 2026

Basophil Activation Test for Investigation of IgE-Mediated Mechanisms in Drug Hypersensitivity
Published on: September 16, 2011
Evaluation of a radioimmunoassay for histamine measurement in biologic fluids
P McBride1, D Bradley, M Kaliner
1Allergic Diseases Section, National Institutes of Allergy and Infectious Diseases, National Institutes of Health, Bethesda, Md.
Abstract:
A new radioimmunoassay for the measurement of histamine in biologic fluids was evaluated. Assay selectivity and specificity were achieved by "succinyl-glycinamide derivatization" of histamine in samples to mimic the immunogen used to generate the monoclonal antibody. The assay exhibits a linear response from 0.1 to 5.0 ng/ml of histamine and the monoclonal antibody used has partial recognition of only N-methylhistamine (other than histamine). With minimal modifications, the assay can accurately measure histamine in plasma, urine, and buffer. Normal ranges for human subjects were established: plasma levels are 0.193 +/- 0.08 ng/ml (n = 40) and urine levels are 20.9 +/- 11.2 micrograms histamine/gm creatinine (n = 10).

