Radiation exposure during cardiac device implantation: Lessons learned from a multicenter registry
Paul Bru1, Antoine Dompnier2, Walid Amara3
1Department of Cardiology, Groupe Hospitalier de la Rochelle Ré Aunis, La Rochelle, France.
Background:
Little data are available about radiation exposure during cardiac electrical device implantation, and no dose reference levels have been published. This multicenter, prospective, observational study assesses patient and staff radiation exposure during cardiac device implantations, and aims at defining dose reference levels.
Methods:
Patient demographic, procedural, and radiation data were obtained for 657 procedures from nine institutions. Physician and staff exposure were measured using real-time dosimeters worn beneath and above lead apron. Statistical analysis included fluoroscopy time (FT), dose-area product (DAP), and DAP adjusted for FT and body mass index.
Results:
Pacemakers and cardioverter defibrillators were implanted in 481 and 176 patients, respectively. Of these, 152 were treated with cardiac resynchronization therapy (CRT). Median FTs were 837s (interquartile range [IQR]: 480-1323), 117s (IQR: 69-209), and 101s (IQR: 58-162), and median DAPs were 1410 (IQR: 807-2601), 150 (IQR: 72-338), and 129 (IQR: 72-332) cGy.cm² for biventricular, dual chamber, and ventricular device implantation, respectively. Dose reference levels correspond to the third quartile values. During CRT, higher exposure was observed with four X-ray systems than with the two newer and customizable ones (adjusted DAP of 0.90 [IQR: 0.26-1.01] and 0.29 [IQR: 0.23-0.39], respectively; P < .001).
Conclusion:
Based on real-life measurements, this multicenter registry provides dose reference levels and may help centers assess radiation exposure. Although biventricular device implantation was responsible for the highest radiation exposure, FT was meaningfully shortened compared to previously reported values. For a same FT, the use of new generators and custom settings has significantly reduced DAP.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...