Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
DETERMINATION OF TOLERANCE LEVELS IN RADIOTHERAPY DOSIMETRY BASED ON STATISTICAL INTERFERENCE
Irena Koniarová1, Lukáš Kotík1
1National Radiation Protection Institute, Bartoškova 28, 14000 Prague, Czech Republic.
Abstract:
The most important dosimetry quantity that is determined at radiotherapy centers is the absorbed dose to water for external beams. Fixed tolerances for absorbed doses measured under reference conditions with an ionization chamber for high-energy photon and electron beams are usually 2 and 3%, respectively, regardless of uncertainties of the input variables and other conditions during evaluation. In reality, this agreement should be evaluated considering the uncertainties of the input variables because they affect the size of the random deviations of the measurements from their true values. The aim of this work was to develop a new approach to evaluate the agreement between measured and reported values based on statistical interference rather than to use fixed tolerance levels. The proposed method considers different scenarios that can occur during the evaluation of agreement. Because the method is described in general, it can be used in all similar situations when partial uncertainties can be established.
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Biological Effects of Radiation
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Bioequivalence Data: Statistical Interpretation
Dosage Regimen: Individualization