Humidity Effect on the Drift Times of the Reactant Ions in Ion Mobility Spectrometry
Zahra Izadi1, Mahmoud Tabrizchi1,2, Helko Borsdorf3
1Department of Chemistry , Isfahan University of Technology , Isfahan , 84156-83111 , Iran.
Abstract:
The effect of moisture content on the drift times of NH4+ and H3O+ reactant ions at different temperatures was experimentally and theoretically studied using an ion mobility spectrometer (IMS). The peak positions of the ions shifted to higher drift times as the humidity of the drift gas increased. The peak displacements were attributed to the consecutive formation of hydrated ion clusters, RI+(H2O). Using chemical equilibrium relations and thermodynamic data derived from DFT calculation, a model was proposed to formulate the change in the drift times as 1/td = 1/tdΘ - βT log[H2O], where β is a constant and T is temperature. [H2O] is the concentration of water in ppm and, tdΘ is the drift time at the standard condition of [H2O] = 1 ppm. The proposed equation perfectly predicted the change in the drift times of the reactant ions as a function of the moisture in the drift gas. Accordingly, standard mobility, K = KΘ - γT ln[H2O], was defined, which is independent of the moisture level of the drift gas and reflects the chemical reactivity. In this work, it is proposed to correct the reported reduced mobilities for the moisture to a standard condition of 1 ppm water concentration, in a similar manner to the corrections to the standard temperature and pressure of 273 K and 760 mbar, respectively. Finally, the likelihood that different hydrates forms of the reactant ion exist is discussed based on the entropy concept.
More Related Videos
16:40T-wave Ion Mobility-mass Spectrometry: Basic Experimental Procedures for Protein Complex Analysis
Published on: July 31, 2010
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Factors Affecting Activity Coefficient
The activity coefficient value for an ion is close to one when the solution has almost zero ionic strength, i.e., when the solution shows close to ideal behavior. As the ionic strength of the solution increases from 0 to 0.1 mol/L, a...
Precipitation of Ions
The equation that describes the equilibrium between solid calcium carbonate and its solvated ions is:
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Measuring Reaction Rates
Common Ion Effect
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...