Testing cholesterol accuracy. Performance of several common laboratory instruments.

D D Koch1, D J Hassemer, D A Wiebe

  • 1Department of Pathology and Laboratory Medicine, University of Wisconsin, Madison 53792.

JAMA
|November 4, 1988
PubMed
Summary

Accurate cholesterol testing is vital for assessing heart disease risk. Several routine laboratory methods meet the National Cholesterol Education Program

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Glucose challenge test screening for prediabetes and undiagnosed diabetes.

Diabetologia·2009
Same author

Severe hypercholesterolemia mediated by lipoprotein X in patients with chronic graft-versus-host disease of the liver.

Bone marrow transplantation·2004
Same author

A prospective multicenter clinical trial to evaluate the safety and effectiveness of the implantable miniature telescope.

American journal of ophthalmology·2004
Same author

Peer review: a group activity.

Journal of cataract and refractive surgery·2001
Same author

Intrastromal corneal ring segments for low myopia: a report by the American Academy of Ophthalmology.

Ophthalmology·2001
Same author

Corneal complications and NSAIDs.

Ophthalmology·2001

Area of Science:

  • Clinical Chemistry
  • Laboratory Medicine
  • Cardiovascular Diagnostics

Background:

  • National Cholesterol Education Program (NCEP) guidelines emphasize accurate cholesterol testing for coronary risk assessment.
  • Current NCEP recommendations include 5% limits for imprecision and bias, with a goal to reduce these to 3% by 1992.
  • Clinicians require awareness of laboratory method performance to ensure reliable cholesterol values.

Purpose of the Study:

  • To evaluate the performance of commonly used enzymatic cholesterol assays against NCEP goals.
  • To determine if current laboratory methods can achieve the stricter 3% precision and accuracy targets.

Main Methods:

  • Patient serum samples were analyzed for cholesterol using multiple commercial enzymatic methods.
  • Comparison was made against the Abell-Kendall reference method.

Related Experiment Videos

  • Method precision was assessed using serum-based controls, with instruments operated per manufacturer instructions.
  • Main Results:

    • The DuPont aca, Kodak Ektachem, Hitachi 737, and Cobas FARA methods demonstrated acceptable cholesterol determination.
    • All evaluated methods met the NCEP performance goals for accuracy and precision.
    • The 3% goals for precision and accuracy are achievable with these routinely used enzymatic methods.

    Conclusions:

    • Several commercially available enzymatic cholesterol assays meet current and upcoming NCEP performance standards.
    • These methods provide reliable cholesterol measurements essential for clinical decision-making in cardiovascular risk assessment.
    • The findings support the achievability of stringent laboratory performance criteria for cholesterol testing.