Post-operative Sore Throat: Comparing the Monitored Endotracheal Tube Cuff Pressure and Pilot Balloon Palpation
Nagappan Ganason1, Vanitha Sivanaser2, Chian Yong Liu1
1Department of Anaesthesiology and Intensive Care, Universiti Kebangsaan Malaysia Medical Centre, Kuala Lumpur, Malaysia.
Background:
Endotracheal tube cuff (ETTc) inflation pressure is usually not regarded as an important aspect during intubation. In this study, we compared measuring ETTc pressure and pilot balloon palpation method in causing post-operative airway complications.
Methods:
Two hundred and ninety-two surgical patients requiring intubation were recruited into this prospective, double-blind, randomised controlled study. Group A patients had their ETTc initially inflated, checked by a cuff pressure gauge, recorded and then set to 25 cmH2O. Group B patients had their ETTc inflated using the pilot balloon palpation method. Patients were then followed up for post-operative sore throat, hoarseness and cough.
Results:
The overall incidence of post-operative sore throat was 39.0% versus 75.3% (P < 0.001), hoarseness 6.2% versus 15.1% (P < 0.05) and cough 7.5% versus 21.9% (P < 0.05) in Group A and B, respectively. Group A patients experienced a significant reduction in the incidence and severity of sore throat up to 24 h post-operatively (P < 0.001), hoarseness at the first hour (P = 0.004) and cough at first and 12 h post-operatively (P = 0.002).
Conclusion:
Adjusting the ETTc pressure to 25 cmH2O reduces post-operative sore throat, hoarseness and cough compared to pilot balloon palpation method.
More Related Videos
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Physical Assessment of the Respiratory Tract II: Palpation
Thoracic Palpation
Thoracic palpation detects tenderness, masses, lesions, respiratory excursions, and vocal fremitus. The nurse assesses...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...