Limitations of ion mobility spectrometry-mass spectrometry for the relative quantification of architectural isomeric
Romain Liénard1,2, Quentin Duez1, Scott M Grayson3
1Interdisciplinary Center for Mass Spectrometry, Organic Synthesis and Mass Spectrometry Laboratory, University of Mons-UMONS, Mons, Belgium.
Abstract:
Since their discovery, cyclic polymers have attracted great interest because of their unique properties. Today, the preparation of these macrocyclic structures still remains a challenge for polymer chemists, and most of the preparation pathways lead to an inescapable contamination by linear by-products. As the properties of the polymers are closely related to their structure, it is of prime importance to be able to assess the architectural purity of a sample.
Methods:
In this work, the suitability of ion mobility spectrometry-mass spectrometry (IMS-MS) for the quantification of two isomers was investigated. A cyclic poly(L-lactide) was prepared through photodimerization of its linear homologue. Since IMS-MS can be used to differentiate cyclic polymer ions from their linear analogues because of their more compact three-dimensional conformation, the present work envisaged the use of IMS-MS for the quantification of residual linear polymers within the cyclic polymer sample.
Results:
Using the standard addition method to plot calibration curves, the fraction of linear contaminants in the sample was determined. By doing so, unrealistically high values of contamination were measured.
Conclusions:
These results were explained by an ionization efficiency issue. This work underlines some intrinsic limitations when using IMS-MS in the context of the relative quantification of isomers having different ionization efficiencies. Nevertheless, the linear-to-cyclic ratio can be roughly estimated by this method.
More Related Videos
09:34Improved Polymerase Chain Reaction-restriction Fragment Length Polymorphism Genotyping of Toxic Pufferfish by Liquid Chromatography/Mass Spectrometry
Published on: September 20, 2016
06:56Characterization of Synthetic Polymers via Matrix Assisted Laser Desorption Ionization Time of Flight MALDI-TOF Mass Spectrometry
Published on: June 10, 2018
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Mass Spectrometry: Isotope Effect
Mass Spectrum: Interpretation
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrometry: Overview
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences