Validity and reliability of the Modified Sphygmomanometer Test with fixed stabilization for clinical measurement of
Brenno B C Silva1, Ana C T Venturato1, Larissa T Aguiar2
1NeuroEixo, Faculdade Ciências Médicas de Minas Gerais, Belo Horizonte, Minas Gerais, Brazil.
Objective:
To investigate the concurrent validity of the Modified Sphygmomanometer Test (MST) with fixed stabilization, compared to the portable dynamometer, and to verify its test-retest and inter-raterreliability.
Methods:
Methodological study. The muscle strength of the following groups was measured: flexors and extensors of the wrist, of the hip, and of the knee and plantar flexors. The Tycos® aneroid sphygmomanometer and the Microfet 2® dynamometer were used. Stabilization was performed using the Mullingan® belt. Descriptive statistics were performed for characterization of the sample. To determine the validity of the MST with fixed stabilization, comparing it with the portable dynamometer, we investigated the correlation between the measurements obtained with the two instruments using the Pearson correlation coefficient. Intraclass Correlation Coefficient (ICC) was used to investigate inter-examiner and test-retest reliability (α = 0.05).
Results:
59 individuals were included (1176 evaluations). A statistically significant correlation of moderate to high magnitude (0.58 ≤ r ≤ 0.81) was observed for concurrent validity of the MST with fixed stabilization versus the portable dynamometer for all muscle groups. Regarding inter-rater reliability, it was observed statistically significant ICC considered excellent to good (0.72 ≤ r ≤ 0.94) for all muscle groups. Regarding test-retest reliability, first evaluator presented good to excellent ICC (0.64 ≤ r ≤ 0.94), while second evaluator presented good to excellent ICC (0.74 ≤ r ≤ 0.96) for all muscle groups.
Conclusion:
The MST with fixed stabilization is valid and reliable for clinical measurement of muscle strength and can overcome previous limitations reported in the literature.
More Related Videos
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
09:44Manual Muscle Testing: A Method of Measuring Extremity Muscle Strength Applied to Critically Ill Patients
Published on: April 12, 2011
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure