Intra-observer and Device-Dependent Inter-observer Reliability of Contrast-Enhanced Ultrasound for Muscle Perfusion
Pierre Kunz1, Sophia Kiesl2, Sascha Groß2
1Center for Orthopedics, Trauma Surgery and Spinal Cord Injury, Ultrasound Centre, HTRG, Heidelberg University Hospital, Heidelberg, Germany; Clinic for Shoulder and Elbow Surgery, Catholic Hospital Mainz, Mainz, Germany.
Abstract:
Muscle perfusion quantification by contrast-enhanced ultrasound (CEUS) may facilitate treatment decisions in musculoskeletal disorders. Translation into clinical routine relies on high intra-observer and inter-observer reliability and transferability between ultrasound devices to enable validation and multicenter studies. This study evaluates these aspects for deltoid muscle perfusion quantification, including possible multicenter study setups. One hundred sixty-six CEUS quantifications were conducted on 42 shoulders. Intra-observer reliability revealed a high intra-class correlation coefficient (ICC, r = 0.91) and low coefficient of variation (CV, 10.28%). Inter-observer reliability revealed an ICC of .84 and a CV of 17.1%, but these values decreased when different ultrasound devices were used (ICC = .60, CV = 18.6%). Re-evaluating subgroups with high sectional plane concordance significantly increased reliability (intra-observer: ICC = .97, CV = 5.49%, inter-observer/same device: ICC = .98, CV = 5.83%, varying devices: ICC = .78, CV = 9.8%). CEUS perfusion quantification of the deltoid seems applicable for multicenter studies, yet pooling different ultrasound devices remains critical. Sectional plane concordance appears to be crucial for reliability and transferability of CEUS muscle perfusion quantifications.