Related Experiment Video
Updated: Jan 3, 2026

Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
Contemporary portable oxygen concentrators and diverse breathing behaviours -- a bench comparison
1ResMed Science Center, ResMed Ltd, Elizabeth Macarthur Drive, Bella Vista, Sydney, Australia. Dion.Martin@resmed.com.au.
Background:
Decades of clinical research into pulsed oxygen delivery has shown variable efficacy between users, and across a user's behaviours (sleep, rest, activity). Modern portable oxygen concentrators (POCs) have been shown as effective as other oxygen delivery devices in many circumstances. However, there are concerns that they are not effective during sleep when the breathing is shallow, and at very high respiratory rates as during physical exertion. It can be challenging to examine the determinants of POC efficacy clinically due to the heterogeneity of lung function within oxygen users, the diversity of user behaviour, and measurement issues. Representative bench testing may help identify key determinants of pulsed-oxygen device efficacy.
Methods:
Three contemporary devices were bench-evaluated across three simulated breathing behaviours: activity, rest, & oronasal breathing during sleep. Emphasis was placed on breathing patterns representative of oxygen users.
Results:
All three POCs performed well during simulated breathing during exertion and at rest. Differences in triggering ability were noted for the scenario of oronasal breathing during sleep.
Conclusions:
The results are supportive of contemporary POC triggering abilities. The differences shown in ultimate trigger sensitivity may have relevance to oronasal breathing during sleep or other challenging scenarios for pulsed oxygen delivery, such as dominant mouth breathing during exertion or unfavourable nasal geometry.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...