Related Experiment Video
Updated: Jan 3, 2026

Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
Fiberoptic-Guided and Blind Tracheal Intubation Through iLTS-D, Ambu® Auragain™, and I-Gel® Supraglottic Airway
Mostafa Somri1, Ibrahim Matter2, Luis A Gaitini1
1Department of Anesthesiology, Bnai Zion Medical Center, Haifa, Israel; Israel Faculty of Medicine, Technion, Israel-Institute of Technology, Haifa, Israel.
Background:
The use of supraglottic airway devices (SADs) is becoming more widespread. However, there is little evidence to show which device is best in an emergent clinical scenario.
Objective:
We compared both fiberoptic-guided and blind tracheal intubation through the Intubating Laryngeal Tube Suction-Disposal (iLTS-D), the AuraGain™, and the i-gel® in an airway manikin.
Methods:
Thirty residents were included in a randomized trial to perform both fiberoptic-guided and blind tracheal intubation using the iLTS-D, the AuraGain, and the i-gel. The main endpoint was the total time taken to achieve successful fiberoptic intubation through the SAD. Additional endpoints included total time for blind intubation, SAD insertion time, tracheal tube insertion time, intubation success rate, fiberoptic view, and maneuvers performed to achieve tracheal intubation.
Results:
All participants performed fiberoptic intubation using all three SADs on the first attempt. The total time to fiberoptic tracheal intubation using the i-gel, AuraGain, and iLTS-D was 42 s, 56 s, and 56 s, respectively. The blind tracheal intubation success rate was 80% with the iLTS-D, 43% with the i-gel, and 0% with the AuraGain. The total time for blind tracheal intubation through the i-gel and the iLTS-D was 29 s and 40 s, respectively. Laryngeal view grades were significantly poorer with the iLTS-D compared to the other devices. The iLTS-D required significantly more maneuvers to achieve successful tracheal intubation.
Conclusions:
In an airway manikin, the iLTS-D, AuraGain, and i-gel appear to be reliable devices for airway rescue and fiberoptic-guided tracheal intubation. The iLTS-D is recommended for blind tracheal intubation.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...