Design of band-pass filters in fluoroscopy by experimental and simulation methods for the 40 keV energy X-ray
Goli Khaleghi1, Jamshid Soltani-Nabipour2, Mahdi Sadeghi3
1Medical Radiation Engineering Department, Science and Research Branch University, P.O. Box: 14515-775, Tehran, Iran.
Abstract:
This study was carried out empirical computational and to design filters that while eliminating low-energy radiations according to the conventional methods attenuate high-energy beams that do not fundamentally affect images quality improvement and the absorbed patient dose reduction as well. In this regard, the impacts of thickness and filter material were examined on the contrast, resolution, absorbed patient dose, and image quality. We found that the use of filters increases the resolution, image quality and reduces the output dose intensity greatly, and the 0.1 mm thickness tin element was selected as the most suitable element for the filter.
Related Concept Videos
Bandpass Sampling
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2....
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Active Filters
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...