Background:
Due to the growing epidemic of atrial fibrillation (AF), new strategies for AF screening, diagnosis, and monitoring are required. Wearable devices with on-board AF detection algorithms may improve early diagnosis and therapy outcomes. In this work, we implemented optimized algorithms for AF detection on a wearable ECG monitoring device and assessed their performance.
Methods:
The signal processing framework was composed of two main modules: 1) a QRS detector based on a finite state machine, and 2) an AF detector based on the Shannon entropy of the symbolic word series obtained from the instantaneous heart rate. The AF detector was optimized off-line by tuning its parameters to reduce the computational burden while preserving detection accuracy. On-board performance was assessed in terms of detection accuracy, memory usage, and computation time.
Results:
The on-board implementation of the QRS detector produced an overall accuracy of 99% on the MIT-BIH Arrhythmia Database, with memory usage = 672 bytes, and computation time ≤90 μs. The on-board implementation of the optimized AF algorithm gave an overall accuracy of 98.1% (versus 98.3% of the original version) on the MIT-BIH AF Database, with increased sensitivity (99.2% versus 98.5%) and decreased specificity (97.3% versus 98.2%), memory usage = 4648 bytes, and computation time ≤ 75 μs (consistent with real-time detection).
Conclusions:
This study demonstrated the feasibility of real-time AF detection on a wearable ECG device. It constitutes a promising step towards the development of novel ECG monitoring systems to tackle the growing AF epidemic.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias V: Evaluating Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...