Analytical effects on clumped isotope thermometry: Comparison of a common sample set analyzed using multiple
William F Defliese1,2, Aradhna Tripati3,4
1Department of Earth, Planetary, and Space Sciences, University of California, Los Angeles, CA, USA.
Rationale:
Carbonate clumped isotope geothermometry is being increasingly used in multiple disciplines in the geosciences. However, potential interlaboratory issues are arising from different standardization procedures that may contribute to the multiple Δ47 -temperature calibrations reported in the literature. We investigate this issue by comparing a common temperature calibration sample set across three different mass spectrometers, using multiple standardization methods.
Methods:
The same temperature calibration sample set was analyzed on three different mass spectrometers. Several standardization methods were utilized, including the use of carbonate versus gas standards, and different types of background correction were applied to the raw data.
Results:
All standardization types applied resulted in statistically indistinguishable Δ47 -temperature slopes, with the exception of standardization calculations that did not correct for background effects. Some instruments and standardizations showed different intercepts relative to each other. The use of carbonate standards improved comparability between different instruments relative to gas standards.
Conclusions:
Our results show that background effects are the largest factor potentially affecting Δ47 results, and there may be an improvement in interlaboratory precision using carbonate standards. Critically, all techniques used for standardizing Δ47 results converge on a common slope as long as background effects are properly corrected. The use of carbonate standards is recommended as a component of standardization procedures.
More Related Videos
06:06A Simple Dewar/Cryostat for Thermally Equilibrating Samples at Known Temperatures for Accurate Cryogenic Luminescence Measurements
Published on: July 19, 2016
10:16Design and Operation of a Continuous 13C and 15N Labeling Chamber for Uniform or Differential, Metabolic and Structural, Plant Isotope Labeling
Published on: January 16, 2014
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
Mass Spectrometry: Isotope Effect
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...