Myocardial deformation and acute cellular rejection after heart transplantation: Impact of inter-vendor variability
Martín Ruiz-Ortiz1, Sara Rodriguez-Diego1, Mónica Delgado1
1Cardiology Department, Reina Sofia University Hospital, Cordoba, Spain.
Insights
Inter-vendor variability in myocardial strain analysis impacts detecting acute cellular rejection (ACR) in heart transplant recipients. Results showed inconsistent associations between strain parameters and ACR across different software, necessitating validation.
Area of Science:
- Cardiology
- Transplant Medicine
- Medical Imaging
Background:
- Acute cellular rejection (ACR) is a significant concern in heart transplant recipients.
- Myocardial strain analysis via echocardiography is a potential non-invasive tool for monitoring graft health.
- Inter-vendor variability in imaging analysis may affect diagnostic accuracy.
Purpose of the Study:
- To assess the impact of inter-vendor variability in myocardial strain analysis for detecting ACR in heart transplant patients.
- To investigate the association between strain parameters and ACR using two different software packages.
Main Methods:
- Serial echocardiograms and endomyocardial biopsies (EMB) were performed in 18 heart transplant recipients within their first year post-transplant.
- Myocardial strain was analyzed using two distinct software packages at different institutions.
- Association between strain values and ACR (any grade or grade ≥2R) was examined.
Main Results:
- Low inter-vendor reproducibility was observed for most strain parameters (left ventricle longitudinal, radial, and right ventricle longitudinal strain).
- Left ventricle circumferential strain (software 1) and radial strain (software 2) showed association with grade ≥2R ACR.
- Right ventricle longitudinal strain (software 1) was associated with ACR of any grade.
Conclusions:
- Inter-vendor reproducibility of myocardial strain values is low, limiting its reliable use in detecting ACR.
- While some strain parameters showed associations with ACR, these findings were inconsistent between the analyzed software.
- Specific validation of echocardiographic analysis software is crucial for accurate ACR detection in heart transplant recipients.
Purpose:
Our objective was to investigate the impact of inter-vendor variability in the ability of myocardial strain analysis to detect acute cellular rejection (ACR) in heart transplant recipients.
Methods:
We performed serial echocardiographic examinations in 18 consecutive adult heart transplanted patients, in their first year post-transplantation, within 3 hours of the routine surveillance endomyocardial biopsies (EMB) in a single center. Myocardial strain was analyzed using two software in two different institutions, and inter-vendor variability of strain values and its association with ACR (any grade or grade ≥2R) was investigated. The parameter of comparison was the peak value of the average curve of strain during the entire cardiac cycle.
Results:
A total of 147 pairs of EMB-echocardiogram were performed, 65 with no ACR, 63 with ACR grade 1R, and 19 with ACR grade ≥2R. Intra-class correlation coefficients for left ventricle longitudinal, radial, and circumferential strain were 0.38, 0.39, and 0.77, respectively, and 0.32 for right ventricular longitudinal strain. Neither software found significant association of left ventricular longitudinal strain with rejection. Grade ≥2R ACR was associated with left ventricular circumferential strain measured with the first software and with left ventricular radial strain with the other; and ACR of any grade was only significantly associated with right ventricle longitudinal strain measured with the first software.
Conclusions:
Inter-vendor reproducibility of strain values was low in this study. Some strain parameters were associated to ACR, although these results were inconsistent between two commercially available software. Specific validation of each software is warranted for this clinical indication.
Related Concept Videos
Myocarditis II: Clinical Features and Diagnostic Tests
Cardiomyopathy V: Interprofessional Care
Tissue Transplantation
The Biology of Tissue Transplantation
The biology of tissue transplantation hinges on the Major Histocompatibility Complex (MHC) molecules. These molecules...