Assessment of a commercially available veterinary blood pressure device used on awake and anesthetized dogs
Objective:
To compare results of a commercially available device for oscillometrically measured blood pressure (OBP) with invasively measured blood pressure (IBP) in awake and anesthetized dogs.
Animals:
19 adult dogs (mean ± SD body weight, 17.8 ± 7.5 kg).
Procedures:
Blood pressures were measured in dogs while they were awake and anesthetized with isoflurane. The OBP was recorded on a thoracic limb, and IBP was simultaneously recorded from the median caudal artery. Agreement between OBP and IBP was evaluated with the Bland-Altman method. Guidelines of the American College of Veterinary Internal Medicine (ACVIM) were used for validation of the oscillometric device.
Results:
In awake dogs, mean bias of the oscillometric device was -11.12 mm Hg (95% limits of agreement [LOA], -61.14 to 38.90 mm Hg) for systolic arterial blood pressure (SAP), 9.39 mm Hg (LOA, -28.26 to 47.04 mm Hg) for diastolic arterial blood pressure (DAP), and -0.85 mm Hg (LOA, -40.54 to 38.84 mm Hg) for mean arterial blood pressure (MAP). In anesthetized dogs, mean bias was -12.27 mm Hg (LOA, -47.36 to 22.82 mm Hg) for SAP, -3.92 mm Hg (LOA, -25.28 to 17.44 mm Hg) for DAP, and -7.89 mm Hg (LOA, -32.31 to 16.53 mm Hg) for MAP. The oscillometric device did not fulfill ACVIM guidelines for the validation of such devices.
Conclusions And Clinical Relevance:
Agreement between OBP and IBP results for awake and anesthetized dogs was poor. The oscillometric blood pressure device did not fulfill ACVIM guidelines for validation. Therefore, clinical use of this device cannot be recommended.
More Related Videos
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
06:51Development of an Algorithm to Perform a Comprehensive Study of Autonomic Dysreflexia in Animals with High Spinal Cord Injury Using a Telemetry Device
Published on: July 29, 2016
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.