Comparison between blood pressure readings using a mercury versus an aneroid sphygmomanometer
Kinaan Farhan1, S Tahira Shah Naqvi1, Syed Asad Hasan Rizvi2
1Department of Medicine, Jinnah Medical and Dental College.
Objective:
For more than a century since its introduction, mercury sphygmomanometer (HgS) had been the mainstay for office measurement of blood pressure (BP). In light of the environmental and health hazards associated with mercury, there is a need to replace it with mercury-free alternatives all over the world. We aimed to validate the widely used aneroid sphygmomanometer (AnS) by comparing its BP readings against BP readings taken with an HgS.
Methods:
We compared the BP readings using AnS vs. HgS on a sample of 300 patients of 18 years or older age admitted to a tertiary care hospital in Karachi, Pakistan.
Results:
The differences between mean HgS and AnS BP readings were found to be statistically significant (P-value <0.01). The mean systolic blood pressure (SBP) readings of the two devices were still significantly correlated (r = 0.989; P < 0.01). Similarly, the mean diastolic blood pressure (DBP) readings were also significantly correlated (r = 0.988; P < 0.01). The aneroid device identified a higher proportion of hypertensive participants compared to the mercury device.
Conclusion:
The difference in the two devices used was found to be significant; however, the readings were correlated with each other. The AnS significantly overestimated BP readings, thereby identifying a higher proportion of hypertensives as compared to the HgS. There is a considerable room for improvement in the accuracy of the AnS, only then an accurate and a well-calibrated AnS could provide an acceptable alternative to the use of the HgS.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.