The hemodynamics of transcatheter aortic valves in transcatheter aortic valves
Hoda Hatoum1, Scott Lilly2, Pablo Maureira3
1Department of Biomedical Engineering, The Ohio State University, Columbus, Ohio.
Background:
The durability of transcatheter aortic valves (TAVs) remains their greatest disadvantage, given that fixed tissue leaflets are not immune to structural degeneration from calcification and thrombosis. Therefore, a second intervention is necessary, especially given that TAV in low-risk patients has shown noninferior outcomes compared with surgery. This study aimed to assess the hemodynamic and turbulent properties of the flow downstream with different TAV-in-TAV configurations, to offer basic hemodynamic guidance for future interventions when currently implanted valves structurally degrade.
Methods:
Six TAV-in-TAV configurations were chosen: 23 mm Evolut-in-26 mm Evolut, 23 mm Evolut-in-23 mm SAPIEN 3, 26 mm Evolut-in-26 mm Evolut, 26 mm Evolut-in-23 mm SAPIEN 3, 23 mm SAPIEN3-in-26 mm Evolut, and 23 mm SAPIEN3-in-23 mm SAPIEN 3. Their hemodynamic performance was assessed in a pulse duplicator for 100 cycles. High-speed imaging and particle image velocimetry were performed to assess turbulence. Effective orifice area (EOA), pinwheeling index (PI), and Reynolds shear stress (RSS) were evaluated.
Results:
The largest mean EOA was obtained with 23 mm SAPIEN-in-26 mm Evolut (2.07 ± 0.06 cm2), and the smallest was obtained with 23 mm Evolut-in-23 mm SAPIEN (1.50 ± 0.04 cm2) (P < .001). The highest mean PI was obtained with SAPIEN-in-SAPIEN (26.5 ± 2.00%), and the lowest was obtained with 26 mm Evolut-in-26 mm Evolut (7.5 ± 1.6%) (P < .01). At peak systole, the least detrimental RSS range was obtained with 23 mm Evolut-in-26 mm Evolut (up to ∼340 Pa), and the most detrimental RSS range was obtained with 23 mm Evolut-in-SAPIEN (∼900 Pa) (P < .01).
Conclusions:
This study shows that best hemodynamic parameters are TAV-specific (implanted and to be implanted). In addition, it shows that RSS levels, which are indicative of turbulence levels and associated with blood damage, are 2- to 3-fold higher after TAV-in-TAV.